Schwarzbeck Radiating Loops

Danh mục: Antenna

Radiating Loops


Catalogue

Yêu cầu báo giá

Schwarzbeck Radiating Loops

Fill out this field
Vui lòng nhập địa chỉ email hợp lệ.
Fill out this field
Fill out this field
Fill out this field

Schwarzbeck Radiating Loops

FESP 5139 – Radiating Loop

Number of turns: 3
Wire diameter: 1 mm Cu
Maximum Coil Current: 12 A (5 min.)
Nominal Coil Current: 8 A continuous
Maximum magnetic Field Strength:(without distance ring, in coil center) 360 A/m (5 min.)
Nominal Magnetic Field Strength with 50 mm distance ring: 85 A/m continuous
Magnetic Fieldstrength, 1 A Coil Current:
(with 50 mm distance ring)
10.61 A/m
Current required for 1 A/m:
(with 50 mm distance ring)
94.25 mA
Medium coil diameter: 100 mm
Spacing coil center to measurement plane: 50 mm
Mechanical Dimensions: 11 x 11 x 21 cm
Connector: N-female
Usable Frequency Range: 0 – 5 (7) MHz
Inductance: 2.3 μH
Weight: 0.47 kg

FESP 5133-9 – Circular screened Coil

Number of turns: 9
Wire diameter: 1.2 mm Cu
Maximum coil current: 11 A (5 min.)
Nominal coil current: 7 A continuous
Max. magn. field strength: 380.4 A/m (5 min.)
Nominal magn. field strength: 242 A/m continuous
Magnetic field strength, 1 A coilcurrent: (with 50 mm distance ring): 34.58 A/m
Current required for 1 A/m:
(with 50 mm distance ring)
28.9 mA
Flux density with 50 mA at 50 mm distance: 127 dBpT
Inner coil diameter: 133 mm
Medium coil diameter: 134.2 mm
Spacing coil center to measurement plane: 50 mm
Mechanical dimensions: 0.16 m x 0.2 m
Connector: N-jack
Usable frequency range: 10 kHz – 3 MHz
Inductance: ~ 20 μH
Resistance: 0.1 Ω
Weight: 0.8 kg
Standard: VG 95377 Part 13

FESP 5132 – Radiating Loop

Number of turns: 20
Wire diameter: 2 mm Cu (AWG 12)
Maximum Coil Current: 20 A (5 min.)
Nominal Coil Current: 10 A continuous
Max. Magn. Field Strength: 1500 A/m (5 min.)
Nominal Magn. Field Strength: 750 A/m continuous
Magn. Field Strength, 1 A Coil Current:
(with 50 mm distance ring)
75.56 A/m
(95 µT)
Current required for 1 A/m:
(with 50 mm distance ring)
13.23 mA
Medium coil diameter: 120 mm
Spacing coil center to measurement plane: 50 mm
Mechanical Dimensions: 0.16 m x 0.25 (0.3) m
Terminals: 4 mm
Usable Frequency Range: 0 – 150 kHz
Resistance: 0.1 Ω
Weight: 1.2 kg
Standard: ISO 11452-8
MIL 461E RS101
EN  55103 5.18.3.2

FESP 5133-1330 – Circular Coil

Medium coil diameter: 126 mm
Medium Coil Distance to Surface: 50 mm
Number of turns: 225
Wire Diameter: 2 mm
Mechanical Dimensions: 190 x 190 x 290 mm
Terminals: 4 mm Laborstecker
Usable Frequency Range: 0 – 20 (50) kHz
Inductance: 6.5 mH
Resistance: 0.5 Ω
Capacitance: 240-300 pF
Resonance Frequency: 120 kHz
Impedance: |Z|=42 Ω @ 1 kHz
Max. Current: 20 A
Max. Fieldstrength at 50 mmspacing: 17 kA/m
Max. magnetic fieldstrength at 1 A coil current in 50 mm distance: 858.2 A/m
Required Current for 1000 A/m in 50 mm distance: 1.165 A
Magnetic Flux Density at 1 A coil current in 50 mm distance: 1.078 mT
180.66 dBpT
Conversion Current-Magnetic Fielstrength: 58.67 dB/m
Weight: 5.3 kg
Standard: Measuring Method SF 01 G acc. to VG95377

FESP 5135 – Circular unshielded Coil

Medium coil diameter: 0.5 m
Number of turns: 20
Turn layers: 1
Mechanical Dimensions: 500 x 500 x 140 mm
Terminals: 4 mm
Usable Frequency Range: 0 – 300 kHz
Inductance: 485 µH
Resistance: 0.8 Ω
Capacitance: 45 pF
Resonance Frequency: ≈1 MHz
Magnetic Fieldstrength at 1 A coil current: 32.016 A/m
30.1 dBA/m
Continuos Current: <5 A
Maximum Current: 7 A (5 min.)
Max. Magnetic fieldstrength: 224 A/m (5 min.)
Fieldstrength (cont. operation): 160 A/m
Weight: 5.2 kg
Spacer: 100 mm
Standard: EN 55103 – 2 A.3.1

 

FESP 5133 – Circular screened Coil

Number of turns: 36
Wire diameter: 1.26 mm Cu
Maximum coil current: 10 A (5 min.)
Nominal coil current: 5 A continuous
Max. magn. field strength: 1385 A/m (5 min.)
Nominal magn. field strength: 700 A/m continuous
Magnetic field strength, 1 A coilcurrent: (with 50 mm distance ring): 138.5 A/m
Current required for 1 A/m:
(with 50 mm distance ring)
7.22 mA
Magnetic field strength, 1 A coil current: (with 100 mm distance ring) 46.88 A/m
Current required for 1 A/m:
(with 100 mm distance ring)
21.3 mA
Flux density with 50 mA at 50 mm distance: 138.8 dBpT
136.8 dBµA/m
Inner coil diameter: 133 mm
Medium coil diameter: 135.5 mm
Spacing coil center to measurement plane: 50 mm / 100 mm
Mechanical dimensions: 0.16 m x 0.25 (0.3) m
Coil terminals: 4 mm Laborstecker (Standard) oder BNC
Shield: Electrostatic
Usable frequency range: 0 – 200 kHz
Inductance: ≈ 300 μH
Resonance frequency: >500 kHz
Resistance: 0.6 Ω
Weight: 1.1 kg
Standard: MIL 461
EN 55103 – 2 A.2.1 b)
EN 55103 – 2 A.4.1

FESP 5210-1 – Square Induction Coil

Frequency range: DC … 10 MHz
Edge length: 1 m +/- 1 %
Number of turns: 1
Max. cont. current: 100 A
Max. current (10 min): 150 A
Max. current (1 min): 200 A
Max. cont. fieldstrength
(coil center):
90 A/m
Inductance: 3.4 µH
Self resonance: 16 MHz
Material / diameter: Aluminium 12 mm
Mounting tube: 22 x 195 mm
Weight: 1.5 kg
Connector: 4 mm wing terminals
Standard:

 

IEC 61000-4-8
IEC 61000-4-9
IEC 61000-4-10

FESP 5410-1 – Square Induction Coil

Frequency range: DC … 10 MHz
Side length: 1 m +/- 1 %
Number of turns: 1
Magnetic field strength @ 1 A coil current: 0.9 A/m
Max. cont. current: 400 A
Max current (10 min): 600 A
Max. cont. field strength (coil center): 360 A/m
Inductance: 2.9 µH
Material / Cross-section: Aluminium / 10 x 25 mm
Mounting tube: 22 x 195 mm
Weight: 4.1 kg
Connector: M16 thread (male)
Standard:

 

IEC 61000-4-8
IEC 61000-4-9
IEC 61000-4-10

 

MFPO 9760 – Current Transformer and Pulse Generator

Power requirement:  230 V / 50/60 Hz / 16 A
Max. cont. current into coil:  400 A
Max current (5 min):  600 A
Max current (max 5 sec.): 1200 A
Max. field strength (coil center, max 5 sec.): >1000 A/m
Dimensions MFPO 9760 (W x D x H): 650 x 800 x 490
Weight: approx. 60 kg
Connector: male threads M16
Standard: IEC 61000-4-8